Detail of Recall

Type of Product:Device
Medsafe Reference:15849
Brand Name:Intuitive Large and Mega Needle Driver Endowrist Instruments
Model:Models: INS-420006, INS-420194, INS-400006, INS-400194
Affected:Manufactured between July 2007 and October 2011
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Intuitive Surgical Inc
Issue:Intuitive Surgical has become aware of the potential for detachment of the jaw insert for Large Needle Drivers and Mega Needle Drivers. If a jaw insert detaches during surgery, this could result in a device fragment falling into the patient.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/12/2013