Detail of Recall

Type of Product:Device
Medsafe Reference:15832
Brand Name:Varian Eclipse Treatment Planning Software
Model:H48
Affected:
Software version:Version: 11 and 13
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Brand Name:Varian ARIA Oncology Information System
Model:HIT
Affected:
Software version:Version: 11 and 13
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Brand Name:Varian ACUITY Radiation Therapy Simulator System
Model:H77
Affected:
Software version:Version: 11 and 13
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems UK Ltd
Issue:A software anomaly has been identified with the Plan Parameters workspace within Eclipse™ Treatment Planning System, ARIA®, and Acuity. When a dose relevant parameter is changed and "Do not clear MUs and Reference Point dose" is selected, the Monitor Units (MUs) may change without the User being aware of the changes.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/12/2013