Detail of Recall

Type of Product:Device
Medsafe Reference:15819
Brand Name:GE X-Ray Systems with CRT Video Monitor Suspensions
Model:Precision 500D, Legacy, Prestilix, Monitrol ,RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII
Affected:All CRT Monitor suspension installed since Jan 1993 with products Precision 500D, Legacy, Prestilix, Monitrol ,RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:It has been identified during several Preventive Maintenance activities that the 4 outer screws that hold the overhead Video Monitor Suspension to the carriage were found missing since installation of the system. There is a potential that your system has not been installed per specified installation requirements. GE Healthcare is not aware of any fall of the overhead monitor suspension; however, missing screws may increase the risk for fall of the monitor suspension. No injury has been reported due to this issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/12/2013