Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15819 |
| Brand Name: | GE X-Ray Systems with CRT Video Monitor Suspensions |
| Model: | Precision 500D, Legacy, Prestilix, Monitrol ,RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII |
| Affected: | All CRT Monitor suspension installed since Jan 1993 with products Precision 500D, Legacy, Prestilix, Monitrol ,RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | It has been identified during several Preventive Maintenance activities that the 4 outer screws that hold the overhead Video Monitor Suspension to the carriage were found missing since installation of the system. There is a potential that your system has not been installed per specified installation requirements. GE Healthcare is not aware of any fall of the overhead monitor suspension; however, missing screws may increase the risk for fall of the monitor suspension. No injury has been reported due to this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/12/2013 |