Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15803 |
| Brand Name: | Ohmeda Infant Warmers (with Blenders) |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Ohmeda Medical Inc |
| Issue: | Oxygen and Air wall or tank inlet fittings on the back panel, were reversed during assembly. As a result, this may potentially have reversed Air/Oxygen gas concentrations; for example, a setting of 100% Oxygen could have an output of 21% Oxygen, and vice versa. The settings of the blender knob will no longer be accurate in the affected units. This only impacts GE Resuscitation Systems with Blender. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/12/2013 |