Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15803
Brand Name:Ohmeda Infant Warmers (with Blenders)
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Ohmeda Medical Inc
Issue:Oxygen and Air wall or tank inlet fittings on the back panel, were reversed during assembly. As a result, this may potentially have reversed Air/Oxygen gas concentrations; for example, a setting of 100% Oxygen could have an output of 21% Oxygen, and vice versa. The settings of the blender knob will no longer be accurate in the affected units. This only impacts GE Resuscitation Systems with Blender.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/12/2013