Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15787
Brand Name:The Binding Site Group Human IgM kit for use on SPA Plus
Model:NK012.S and NK012.10S
Affected:all lot numbers
Software version:
Recalling Organisation:In Vitro Technologies
26-30 Vestey Drive
Mount Wellington
Panmure
Auckland
Contact Information:0800 468 487
Manufacturer:The Binding Site Group Ltd
Issue:Manufacturer investigations identified that the affected product may be subject to antigen excess effect; even though it's true value was below the claimed level of 56.4g/L. The antigen excess capacity of this product has been reviewed and has been confirmed as correct. Investigations have suggested that the tested sample contained predominantly monoclonal IgM and the manufacturer believes that this particular type of IgM reacted aberrantly with the antibody and subsequently yielded the spurious result.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/12/2013