Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15782
Brand Name:Carestream DRX Evolution with Full Featured Wall Stand
Model:DRX Evolution
Affected:
Software version:
Recalling Organisation:Carestream Health New Zealand Ltd
c/- Quigg Partners
Level 7, 28 Brandon St
Wellington
Contact Information:021 977 215
Manufacturer:Carestream Health Inc (HCIS)
Brand Name:Kodak DirectView DR7500 System with Full Featured Wall Stand
Model:DR7500
Affected:
Software version:
Recalling Organisation:Carestream Health New Zealand Ltd
c/- Quigg Partners
Level 7, 28 Brandon St
Wellington
Contact Information:021 977 215
Manufacturer:Carestream Health Inc (HCIS)
Issue:The manufacturer has received 3 reports (one in Australia and two in the United States) where the wall stand (bucky) continued its downward movement despite the radiographer/operator releasing the motorised WS keypad button which controls the downward movement.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/12/2013