Detail of Recall

Type of Product:Device
Medsafe Reference:15779
Brand Name:Elekta Xio RTP System
Model:MUJ
Affected:
Software version:Versions 4.50 and higher
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:When computing the Effective Depth (with bolus) to the weight point, XiO could use the bolus density for a particular segment of the weight fan when it should be using the CT density. If the CT density is different than the bolus density, the effective depth will be incorrect which will lead to incorrect dose.
A resolution for this problem will be available in Patch Release 4.80.03 and Release 5.00.01
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:2/12/2013