Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15779 |
| Brand Name: | Elekta Xio RTP System |
| Model: | MUJ |
| Affected: | |
| Software version: | Versions 4.50 and higher |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | When computing the Effective Depth (with bolus) to the weight point, XiO could use the bolus density for a particular segment of the weight fan when it should be using the CT density. If the CT density is different than the bolus density, the effective depth will be incorrect which will lead to incorrect dose. A resolution for this problem will be available in Patch Release 4.80.03 and Release 5.00.01 |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 2/12/2013 |