Detail of Recall

Type of Product:Device
Medsafe Reference:15774
Brand Name:Beckman Coulter UniCel DxH 600 & DxH 800 Cellular Analysis Systems
Model:Part Numbers: B23858 & 629029
Affected:
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Brand Name:Beckman Coulter DxH Cleaner
Model:Part Numbers: 628022, 628023
Affected:12 specific lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:Complaints related to optical degradation have increased on the DXH 800 and DXH 600 Systems. Customers have reported that they are experiencing an increase in one or more of the following:
* Flow cell clogs (: : ::)
* LS offset errors during Daily Check or control analysis
* High coefficients of variation (CV) for Latron CP-X Controls for Low angle Light Scatter (LALS) and/or Axial Light Loss (ALL) parameters
* System messages and R-flagged results for NRBC, Diff and Retic during control or patient analysis, such as partial clogs (Low Event Rated with Diff Rs) or a voltage failure (System Event: D with Diff Rs)
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/11/2013