Detail of Recall

Type of Product:Device
Medsafe Reference:15753
Brand Name:Intuitive Surgical da Vinci Patient Side Manipulators
Model:da Vinci S, Si and Si-e Patient Side cart
Affected:
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Intuitive Surgical Inc
Issue:Manufacturer has become aware that friction within certain instrument arms can interrupt smooth instrument motion. This interruption is felt by by the surgeon as resistance in the movement of the master. In this situation the instrument can stall momentarily and then suddenly catch-up to the master psition id the surgeon pushes through the resistance. This resistance is most noticeable when moving the instruments at very slow speeds and during small movements of the masters. This system behaviour is not autonomous instrument motion.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/11/2013