Detail of Recall

Type of Product:Device
Medsafe Reference:15750
Brand Name:Leica M525 F50/C50 Surgical Microscope with XY-Unit
Model:M525 F50 & M525 C50 with XY units
Affected:
Software version:Software versions prior to Pack D
Recalling Organisation:Pacific Dental Specialties Limited
3 Clyde Road
Napier Hill
Napier 4110
Contact Information:021 264 9070
Manufacturer:Leica Microsystems (Schweiz) AG
Issue:Leica Microsystems (Schweiz) AG has discovered that in certain activations of the XY, the system can continue to run laterally to its limit (run-on), but will stop if the surgeon activates any button.
Run-on of the XY can only occur in the rare situation of rapid consecutive activation on the 45°, 135°, 225°, or 315°. In addition, the run-on will stop if any other activation of the XY occurs.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/12/2013