Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15721 |
| Brand Name: | Hospira GemStar Docking station |
| Model: | List Numbers: 13075-XX-07 and 13075-XX-09 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Hospira NZ Ltd 58 Richard Pearse Drive Airport Oaks Mangere Auckland 2022 |
| Contact Information: | 04 384 7463 |
| Manufacturer: | Hospira Inc |
| Issue: | Hospira Inc. has become aware of two situations that may occur when using the GemStar Docking Station list numbers 13075-XX-07 and 13075-XX-09 in conjunction with the GemStar infusion pump. When the docking station is used in conjunction with a GemStar Phase 3 Infusion pump (list numbers 13000-XX, 13100-XX or 13150-XX), the potential exists for the pump to fail to power up while connected to the Docking Station. When a GemStar Phase 4 Infusion pump (list numbers 13086-XX, 13087-XX or 13088-XX) is used in conjunction with both a Docking Station and an External Battery Pack accessory (List 13073-XX), there exists a possibility that the GemStar pump will display error code 11/003 indicating excessive input voltage from the external sources. The 11/003 error code will stop infusion if the error occurs when an infusion is in process. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/11/2013 |