Detail of Recall

Type of Product:Device
Medsafe Reference:15721
Brand Name:Hospira GemStar Docking station
Model:List Numbers: 13075-XX-07 and 13075-XX-09
Affected:
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:Hospira Inc. has become aware of two situations that may occur when using the GemStar Docking Station list numbers 13075-XX-07 and 13075-XX-09 in conjunction with the GemStar infusion pump.
When the docking station is used in conjunction with a GemStar Phase 3 Infusion pump (list numbers 13000-XX, 13100-XX or 13150-XX), the potential exists for the pump to fail to power up while connected to the Docking Station.
When a GemStar Phase 4 Infusion pump (list numbers 13086-XX, 13087-XX or 13088-XX) is used in conjunction with both a Docking Station and an External Battery Pack accessory (List 13073-XX), there exists a possibility that the GemStar pump will display error code 11/003 indicating excessive input voltage from the external sources. The 11/003 error code will stop infusion if the error occurs when an infusion is in process.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/11/2013