Type of Product:Device
Medsafe Reference:15711
Brand Name:ASP Sterrad 100NX Sterilisation System
Model:Product Code: 10104-002
Affected:Manufactured between Oct-2012 and Jun-2013
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:The manufacturer, ASP has determined welds could loosen on the frame of STERRAD® 100NX sterilizers manufactured between October 2012 and June 2013. These welds are used to secure support rails to the top tray assembly for added rigidity. The failure of the broken weld could potentially cause the chassis to deflect concaving downward.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/11/2013