Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15711 |
| Brand Name: | ASP Sterrad 100NX Sterilisation System |
| Model: | Product Code: 10104-002 |
| Affected: | Manufactured between Oct-2012 and Jun-2013 |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | The manufacturer, ASP has determined welds could loosen on the frame of STERRAD® 100NX sterilizers manufactured between October 2012 and June 2013. These welds are used to secure support rails to the top tray assembly for added rigidity. The failure of the broken weld could potentially cause the chassis to deflect concaving downward. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/11/2013 |