| Type of Product: | Device |
| Medsafe Reference: | 15709 |
| Brand Name: | GE Patient Data Module (PDM) |
| Model: | Product Codes: SA3 and SG5 |
| Affected: | Manufactured between Apr-2007 and Dec-2012 |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | The spring inside the mounting latch can become dislodged causing the PDM not to latch properly resulting in an unexpected loss of patient monitoring function. This may also result in the PDM falling onto a patient or caregiver. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/11/2013 |