Detail of Recall

Type of Product:Device
Medsafe Reference:15702
Brand Name:Draeger Monitor AC3000 Spring Arm
Model:G92756 and G94499
Affected:
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:Manufacturer has become aware of a case in which the monitor or support bracket has detached from the joint pin at the bottom of the height adjustable AC3000 Spring Arm and fallen down together with the monitor until it was only held by the electrical cable connections. No one was injured but a falling monitor support has the potential to cause personal injury if the event was repeated. An inspection of a further seven monitor brackets in the same hospital identified severe wear on the ring groove responsible for the joint pin.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/11/2013