Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15702 |
| Brand Name: | Draeger Monitor AC3000 Spring Arm |
| Model: | G92756 and G94499 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | Manufacturer has become aware of a case in which the monitor or support bracket has detached from the joint pin at the bottom of the height adjustable AC3000 Spring Arm and fallen down together with the monitor until it was only held by the electrical cable connections. No one was injured but a falling monitor support has the potential to cause personal injury if the event was repeated. An inspection of a further seven monitor brackets in the same hospital identified severe wear on the ring groove responsible for the joint pin. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/11/2013 |