Detail of Recall

Type of Product:Device
Medsafe Reference:15700
Brand Name:Abbott RealTime High Risk HPV and RealTime High Risk (HR) HPV Application CD-ROM (for m2000 System)
Model:List Numbers: 02N09 & 4N05-01
Affected:Version 1.0
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Molecular Inc
Issue:m2000rt instrument noise is defined as an unexpected cycle-to-cycle variability in the flurorescence signals during the amplification and detection process. Abbott Molecular has identified that the instrument noise lower limit validy parameter for the RealTime HR HPV assay is set outside of the optinal setting to flag excess variability in fluroescence signal. Therefore, in the presence of excess noise, incorrect results could be generated.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/11/2013