Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15699
Brand Name:Vitros Chemistry products K+ Slides, Coatings 0884 amd 0885
Model:8157596
Affected:Coatings: 0884, 0885
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics c/- Johnson & Johnson
13A Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 800 705
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:Ortho Clinical Diagnostics (OCD) determined that VITROS® Chemistry Products K+ Slides, Coatings 0884 and 0885 contain a surfactant that was unintentionally added during the manufacturing process. OCD has conducted testing that indicates that VITROS® K+ Slides made from Coatings 0884 and 0885 are currently meeting their intended use and performance specifications. However, at this time we are unable to determine the effect that the addition of the surfactant may have on long-term product performance.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/11/2013