Type of Product:Device
Medsafe Reference:15666
Brand Name:Boston Scientific Vessix generator System
Model:RDBS010
Affected:
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Vessix Vascular Inc
Issue:The manufacturer is initiating a Recall of the Vessix™ Generator System. An investigation has found that the units manufactured since 2013-04 (April 2013) were not manufactured according to specification. This manufacturing issue may cause an electrical short in the generator. To date, there have been no reports of patient injury. While there are no reports of patient injury, the issue could result in excessive RF energy being delivered to the renal artery. The potential health hazards associated with the delivery of excessive RF energy include thrombosis formation, nerve damage and thermal damage to the renal artery or adjacent structures.
This Recall affects the generator only and does not impact the Vessix™ Renal Denervation Catheters or any other components of the system. .
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory