Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15655 |
| Brand Name: | Draeger Infinity Acute Care System (IACS) in Neonatal Mode |
| Model: | |
| Affected: | |
| Software version: | VG2.03 & higher |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | The manufacturer has determined that a low SpO2 (yellow) medium grade alarm which is pased by the user using teh alarm pause function, may cause the central station monitoring the device not to alarm if the SpO2 level drops to high grade life threatening (red) alarm. The device itself will alarm. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/11/2013 |