Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15654 |
| Brand Name: | Siemens BFT II Analyser |
| Model: | Catalogue Number: OVKF03 / SMN 10458677 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Products GmbH |
| Issue: | It has been confirmed that Standard Human Plasma (SHP, REF ORKL) pre-dilution ratios given in the Reference Guides of the BFT* II Analyzer do not correspond to the standard dilution procedure for endogenous coagulation factor determination. The incorrect dilution ratio stated in the Reference Guides is limited to the lowest calibration point. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/11/2013 |