Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15653 |
| Brand Name: | Siemens Artis zee Ceiling Systems |
| Model: | Catalogue number – 10094137 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | Cables in the cable harness inside the ceiling carriage may have received wear and tear or damage. In versions before a modification in 2009, a protective sleeve was covering the total length of cabling inside the ceiling carriage. This modification was made in a way that the protective sleeve is no longer covering all cables in total length inside the ceiling carriage. There is now friction between individual cables causing stress over a period of time which for smaller dimensioned cables may result in tearing and system malfunction. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/11/2013 |