Detail of Recall

Type of Product:Device
Medsafe Reference:15653
Brand Name:Siemens Artis zee Ceiling Systems
Model:Catalogue number – 10094137
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Cables in the cable harness inside the ceiling carriage may have received wear and tear or damage.
In versions before a modification in 2009, a protective sleeve was covering the total length of cabling inside the ceiling carriage. This modification was made in a way that the protective sleeve is no longer covering all cables in total length inside the ceiling carriage. There is now friction between individual cables causing stress over a period of time which for smaller dimensioned cables may result in tearing and system malfunction.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/11/2013