| Type of Product: | Device |
| Medsafe Reference: | 15645 |
| Brand Name: | Hill-Rom Clin Activ and Clin Activ Plus Therapy Surface |
| Model: | N02050, N02051, P02062, P02063, P02064, P02065 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Hill-Rom SAS |
| Issue: | Manufacturer has become aware of a possible hazardous situation when silicon oil contained in the Vario sensor within the ClinActiv Therapy Surface is accidentally released. When the pocket of silicone oil is subjected to excessive mechanical stress, it may disconnect and leak oil outside the mattress enclosure and onto the floor. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/12/2013 |