Detail of Recall

Type of Product:Device
Medsafe Reference:15645
Brand Name:Hill-Rom Clin Activ and Clin Activ Plus Therapy Surface
Model:N02050, N02051, P02062, P02063, P02064, P02065
Affected:All lots
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Hill-Rom SAS
Issue:Manufacturer has become aware of a possible hazardous situation when silicon oil contained in the Vario sensor within the ClinActiv Therapy Surface is accidentally released. When the pocket of silicone oil is subjected to excessive mechanical stress, it may disconnect and leak oil outside the mattress enclosure and onto the floor.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/12/2013