Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15637
Brand Name:Smiths Medical Epidural Minipack Systems
Model:100/391/116
Affected:All lots
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical International Ltd (UK)
Issue:The wording in the current IFU states : "If resistance is felt upon withdrawal, consult current medical literature for specific techniques". This is to be replaced by : "If resistance is felt or if the catheter stretches excessively on withdrawal, cease removal. Never exert excessive force as this may compromise the integrity of the catheter. If clinically appropriate, reposition the patient (flex or extend, sit up or lay down) and slowly withdraw the catheter. If further resistance, repeat step and/or allow the patient to relax for several minutes/hours and attempt to remove later. If unable to remove, consult anesthetist. After removal, examine the catheter to ensure that the entire catheter with tip is intact. Should the tip not be intact, consult the anesthetist"
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/11/2013