Detail of Recall

Type of Product:Device
Medsafe Reference:15630
Brand Name:GE CARESCAPE Monitor B850
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Issue:The Patient data Module (PDM) can sometimes spontaneosuly lose the communication withthe host patient monitor when atached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware.Communication is not automatically re-established. Loss of communication with the host monitor results in a loss of monitoring for the parameters provided by the PDM module.. The parameters may include ECG, SpO2, NIBP, CO, Temperature and Respiration.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/11/2013