Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15630 |
| Brand Name: | GE CARESCAPE Monitor B850 |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Finland Oy |
| Issue: | The Patient data Module (PDM) can sometimes spontaneosuly lose the communication withthe host patient monitor when atached to the F5-01 Frame due to an issue with the PDM Ethernet communication hardware.Communication is not automatically re-established. Loss of communication with the host monitor results in a loss of monitoring for the parameters provided by the PDM module.. The parameters may include ECG, SpO2, NIBP, CO, Temperature and Respiration. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/11/2013 |