Detail of Recall

Type of Product:Device
Medsafe Reference:15629
Brand Name:Welch Allyn Propaq LT/ Propaq 802 Patient Monitor
Model:802LTAN, 802LTON, 802LTRN
Affected:
Software version:
Recalling Organisation:Medxus Ltd
Unit 4
100 Fitgerald Avenue
Christchurch 8011
Contact Information:03 365 7403
Manufacturer:Welch Allyn Protocol Inc
Issue:Internal serial testing has identified fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity that could, in rare instances, potentially interrupt PROPAQ LT device operations. If this were to occur the device would either present an error screen instructing the user to restart the monitor, or could present a blank screen or white screen.Either screen would be accompanied by one or more blimking LED lights, and in either case, the isue would be resolved by restarting (power cycling) the monitor.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/11/2013