Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15623 |
| Brand Name: | Radiometer ABL90 Flex analyser |
| Model: | |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | RADIOMETER has recently become aware of a potential issue of negative bias on Sodium results if the ABL90 analyser is not on a horizontal table. The error has only been reported when the analyzer has been installed on a roll stand and tilted (i.e. the support plate is not horizontal) and the analyser has been idle for more than 30 minutes. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/11/2013 |