Detail of Recall

Type of Product:Device
Medsafe Reference:15623
Brand Name:Radiometer ABL90 Flex analyser
Model:
Affected:All serial numbers
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:RADIOMETER has recently become aware of a potential issue of negative bias on Sodium results if the ABL90 analyser is not on a horizontal table. The error has only been reported when the analyzer has been installed on a roll stand and tilted (i.e. the support plate is not horizontal) and the analyser has been idle for more than 30 minutes.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/11/2013