Type of Product:Device
Medsafe Reference:15617
Brand Name:Abbott i-Stat PT/INR Cartridge
Model:Model: 3P89-24
Affected:Lots from C13139 to C13270A
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:Abbott Point of Care Canada Ltd
Issue:The manufacturer has determined that a range of i-STAT PT/INR cartridge lots have the potential to exhibit incorrectly elevated results. Internal studies have demonstrated that i-STAT PT/INR results are eleveated by an average of approxiamtely 10% as compated to the International Reference Preparation (rTF/09) in the therapeutic range of 1.8 to 3.0 INR.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/11/2013