Detail of Recall

Type of Product:Device
Medsafe Reference:15596
Brand Name:ASP Sterrad Booster
Model:Catalogue Numbers: 15400
Affected:All units
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Brand Name:ASP Sterrad Adapters
Model:Product codes: 15401, 15402, 15403, 15407
Affected:All units
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:Issue 1: Manufacturer has identified that the IFU for the STERRAD Booster and STERRAD Adaptor include incorrect instructions and potential consequences are failing to use the device when needed (potentially resulting in a lumen which is not sterile) or using the devices when not needed (potentially resulting in residual hydrogen peroxide left in the lumen)
Issue 2: The manufacturer has introduced a new verification tool that should be used with STERRAD® Boosters and STERRAD® Adaptors. This new tool is a checklist that should be printed and used to provide verification of the proper use of the STERRAD® Booster/Adaptor before and after the STERRAD® sterilization cycle.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/11/2013