Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15595
Brand Name:Steris Reliance EPS endoscope processor
Model:All models
Affected:All serial numbers
Software version:
Recalling Organisation:Device Technologies New Zealand
47 Arrenway Drive
Albany
AUCKLAND 0632
Contact Information:09 913 2000
Manufacturer:Steris Corporation
Issue:Currently the air pipe on specific Fujinon endoscopes doesn't receive forced flow during the processing cycle. A new STERIS Flow unit 7 has been developed to facilitate forced flow of high level disinfecting solution through the air pipe and is required for Reliance EPS processing specific FUJINON G5 and G8 type flexible GI endoscopes.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/10/2013