Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15590
Brand Name:Microgenics CEDIA Tobramycin II Assay
Model:Catalogue Number: A96842
Affected:Lot: 60169764
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Microgenics Corporation
Issue:In-house testing of the CEDIA® Tobramycin II Assay (LN: 60169764) shows that the reconstituted reagents do not maintain 30 days reconstituted stability. The instability of the reconstituted reagents may cause higher recovery of samples. Hence, lot number 60169764 reagents must be used within 14 days after reconstitution. Reconstituted reagents beyond the 14 days should be discontinued and discarded.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/10/2013