Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15585 |
| Brand Name: | Getinge 46 Series Washer Disinfector with PACS 300 module |
| Model: | |
| Affected: | Manufactured between 2004 and 2008 |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Getinge Disinfection AB |
| Issue: | Investigations have shown that the overheating protection system of the heating element may fail to function under certain circumstances. This could result in a serious injury such as a burn or respiratory complications as a result of inhaling smoke. The reported overheating events have been confined to the machines. Therefore the probability of occurrence of harm to person based on available evidence is considered remote. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/11/2013 |