Detail of Recall

Type of Product:Device
Medsafe Reference:15582
Brand Name:GE Carescape Patient Monitor B650
Model:Model: B650
Affected:S/W versions 1.1.12 and earlier
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Brand Name:GE Carescape Patient Monitor B850
Model:Model: B850
Affected:S/W version 1.0.12 and earlier
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:The manufacturer has determined that selected software versions of B650 and B850 patient monitors have a number of safety related issues including:
1. Non-invasive blood pressure (NIBP) auto-cycling will not start
2. NIBP auto-cycling stops when selecting Audio & Display off
3. NIBP auto-cycling stops when loading data and measuirng NIBP at the same time
4. Lost NIBP measurements
5. NIBP limit values for infants change when loading a new patient data from a TRAM module
6. Wrong ECG bandwith in E-PSM(P) or E-(P)RE(S)TN modules
7. Changing the primary ECG lead selection in the CIC Pro Central Station will not change the primary measurement lead in PDM or TRAM measurement modules
8. A temp hidden alarm message will not re-activate if a different alarm is acknowledged at iCentral
9. Missing audible alarms from remote monitors after broken network connection
10. Missing ECG waveforms on iCentral.
11. iCentral does not notice a monitor exiting the Standby mode
+ 14 other issues
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/11/2013