Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15577 |
| Brand Name: | Medifab Multistand 2000 standing frame |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medifab Limited 1 Pereita Drive Rolleston 7675 |
| Contact Information: | 0800 543 343 |
| Manufacturer: | Timoteos Oy |
| Issue: | The issues is that the size 3 frame was not tested with a particular electric actuator, which, although sufficient on the smaller size one & two frames, has since been found to have failed under use with the larger frame. Potential upgrade with a stronger actuator to prevent a fault occuring on the Multistand 2000. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/10/2013 |