Detail of Recall

Type of Product:Device
Medsafe Reference:15577
Brand Name:Medifab Multistand 2000 standing frame
Model:
Affected:
Software version:
Recalling Organisation:Medifab Limited
1 Pereita Drive
Rolleston 7675
Contact Information:0800 543 343
Manufacturer:Timoteos Oy
Issue:The issues is that the size 3 frame was not tested with a particular electric actuator, which, although sufficient on the smaller size one & two frames, has since been found to have failed under use with the larger frame.
Potential upgrade with a stronger actuator to prevent a fault occuring on the Multistand 2000.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/10/2013