Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15572 |
| Brand Name: | RaySearch RayStation treatment planning system |
| Model: | |
| Affected: | |
| Software version: | 3.0. 3.5 and 4.0 |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | An issue has been found with the dose calculaton for Elekta motorized wedges in RayStaion 3.0. 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees. The error is that the modified energy spectrum has been used not only for the wedged part of the beam but also for the open part of the wedged field. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/10/2013 |