Detail of Recall

Type of Product:Device
Medsafe Reference:15572
Brand Name:RaySearch RayStation treatment planning system
Model:
Affected:
Software version:3.0. 3.5 and 4.0
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:An issue has been found with the dose calculaton for Elekta motorized wedges in RayStaion 3.0. 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees. The error is that the modified energy spectrum has been used not only for the wedged part of the beam but also for the open part of the wedged field.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/10/2013