Detail of Recall

Type of Product:Device
Medsafe Reference:15560
Brand Name:GE Brivo CT315/325, Brivo CT385, Optima CT520, Optima CT540 Systems
Model:Models: Brivo CT315/325, Brivo CT385, Optima CT520, Optima CT540
Affected:All serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Hangwei Medical Systems Co Ltd
Issue:There is an opportunity, while following a specific workflow, to create a film image with one patient's images and another patient's name in the footer. There is also an issue when within Manual Film Composer, if a color image is selected for printing, another patient's black and white image appears in the preview screen and the color image cannot be printed.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/10/2013