Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15544
Brand Name:Carl Zeiss Master 500
Model:000000-1692-983; 1789-811, 1322-734, 1262-306, 1262-307, 9004-00, 1064-064 and 1134-078
Affected:
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:Carl Zeiss Meditec AG
Issue:The ULIB database delivered with new devices as well as the ULIB database offered between March 25 and Sept 24 2013 for download contained an erroneous value for one lens. This could lead to the implantation of a lens with a wrong diopter value.
Specifically the a0 constant of the lens AMO Sensar AS40E is listed as 0.0 in the internal lens data base of the IOL Master. The correct value is -2.420.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/10/2013