Detail of Recall

Type of Product:Device
Medsafe Reference:15506
Brand Name:Biomerieux Vitek MS Instrument
Model:
Affected:All instruments with a Serial Number lower than #50039, Serial Number #50050, or Serial Number greater than #60000
Software version:
Recalling Organisation:BioMerieux New Zealand Ltd
Suite 4, The Business Connection Office
2 Kalmia Street
Auckland 1544
Contact Information:0800 284 825
Manufacturer:Kratos Analytical Ltd
Issue:The VITEK MS instrument is equipped with a Transient Recorder (TR) Board associated with a specific firmware. This association is used to compute the time of flight of detected ions. Some issues have been reported when firmware version 3.13 was installed on systems equipped with the old TR Board (ref. 6200350). Extensive testing has proven that firmware 3.13 may potentially cause an instrument to freeze or hang during acquisition when installed with the old TR board, with a risk of repeated results for different spots.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/10/2014