Type of Product:Device
Medsafe Reference:15494
Brand Name:Siemens Dimension Vista 400 & 1500 UDM Flex Assignment
Model:SMN 10488224 & 10444802
Affected:Version: Vista SW Version 3.5.1 and below
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Siemens Healthcare Diagnostics has confirmed that under a rare set of conditions when utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex cartridge to process the user defined method. Investigation of the issue has determined that this error will only occur if the Assign Empty Reagent Cartridge window has been left open and the Flex inventory on the system changes.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/10/2013