Detail of Recall

Type of Product:Device
Medsafe Reference:15482
Brand Name:GE Avance, Amingo, Avamce and Aisys Anaesthesia Work Stations
Model:
Affected:
Software version:8.00 and 10.00
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:While operating in Presure Control Ventilation - Volume Guarantee (PCV-VG) mode, the affected products may produce an over delivery tidal volume when given a unique sequence of user inputs and a collapsed bellows (partially filled circuit).Exposure to excessive tidal volumes while pressure limits are maintained can result in pulmonary volutrauma and/or an increased pulmonary vascular resistance that can negatively affect cardiac filling pressure and/or stroke volume.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/10/2013