Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15481 |
| Brand Name: | Siemens Ysio |
| Model: | 10094200.10094910, 10281163, 10281013 & 10502200 |
| Affected: | |
| Software version: | VC10 VB10 |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Brand Name: | Siemens Luminos Agile & dRF |
| Model: | 10094200.10094910, 10281163, 10281013 & 10502200 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Brand Name: | Siemens Uroskop Omnia |
| Model: | VC10 |
| Affected: | 10094200.10094910, 10281163, 10281013 & 10502200 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | During a RAD examination with automatic exposure control, a highly unlikely malfunction can result in radiation continung longer than required.This may also result in an overexposed image that is not of diagnostic quality, so that th examination needs to be repeated. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/10/2013 |