Type of Product:Device
Medsafe Reference:15471
Brand Name:GE Patient Monitor Multiparameter
Model:Models: B20 V1, B30, B40 V1, B40 V2
Affected:Selected serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:GE Healthcare has recently become aware of a potential safety issue due to ECG filter settings associated with the B20/B40 V1, B30 and B40 V2 Patient Monitors. The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly if the monitor starts with impedance respiration set in the ON position. Instead of the user selected filter, an additional 0.5 Hz high pass filter is used. The additional 0.5 Hz high pass filter causes changes to the ECG waveform morphology and incorrect display of ST values. The ST-elevation measurement can be under estimated in this situation and therefore the treatment of the patient can be delayed.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/10/2013