Detail of Recall

Type of Product:Device
Medsafe Reference:15448
Brand Name:Beckman Coulter UniCell DxH 800 Cellular Analysis system
Model:Product Number: 629029
Affected:Selected serial numbers
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Brand Name:Beckman Coulter UniCel DxH 600 Cellular Analysis System
Model:Product Code: B23858
Affected:Selected serial numbers
Software version:
Recalling Organisation:Beckman Coulter Australia Pty Ltd
1 Dale Street
Brookvale
NSW 2100
Australia
Contact Information:
Manufacturer:Beckman Coulter Inc
Issue:An internal investigation indicates that specific lots of check valves used in several locations with the LH750 and LH780 analysers may fail. This failure may result in an air or liquid leak at the VCS Waste Chamber or Probe Waste Chamber.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/9/2013