Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15448 |
| Brand Name: | Beckman Coulter UniCell DxH 800 Cellular Analysis system |
| Model: | Product Number: 629029 |
| Affected: | Selected serial numbers |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Brand Name: | Beckman Coulter UniCel DxH 600 Cellular Analysis System |
| Model: | Product Code: B23858 |
| Affected: | Selected serial numbers |
| Software version: | |
| Recalling Organisation: | Beckman Coulter Australia Pty Ltd 1 Dale Street Brookvale NSW 2100 Australia |
| Contact Information: | |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | An internal investigation indicates that specific lots of check valves used in several locations with the LH750 and LH780 analysers may fail. This failure may result in an air or liquid leak at the VCS Waste Chamber or Probe Waste Chamber. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/9/2013 |