Detail of Recall

Type of Product:Device
Medsafe Reference:15445
Brand Name:Beckman Coulter LH 750 Hematology Analyser
Model:Part Numbers: 6605632 & A85570
Affected:Multiple serial numbers
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Brand Name:Beckman Coulter LH 780 Hematology Analyser
Model:Product Numbers: 723585 & A90728
Affected:Selected serial numbers
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:An internal investigation indicates that specific lots of check valves used in several locations with the LH750 and LH780 analysers may fail. This failure may result in an air or liquid leak at the connection to the Vacuum Overflow Tank waste line or at the Backwash Tank drain line.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/9/2013