Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15445 |
| Brand Name: | Beckman Coulter LH 750 Hematology Analyser |
| Model: | Part Numbers: 6605632 & A85570 |
| Affected: | Multiple serial numbers |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Brand Name: | Beckman Coulter LH 780 Hematology Analyser |
| Model: | Product Numbers: 723585 & A90728 |
| Affected: | Selected serial numbers |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | An internal investigation indicates that specific lots of check valves used in several locations with the LH750 and LH780 analysers may fail. This failure may result in an air or liquid leak at the connection to the Vacuum Overflow Tank waste line or at the Backwash Tank drain line. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/9/2013 |