Detail of Recall

Type of Product:Device
Medsafe Reference:15377
Brand Name:Kendall Argyle Single Lumen Polyurethane Umbilical Vessel Catheter
Model:Product codes: 8888-160325, 8888-160333, 8888-160341
Affected:All lots
Software version:
Recalling Organisation:Covidien NZ Ltd
Level 3, Building 5, Central Park Corporate Centre
666 Great South Road
Penrose
AUCKLAND 1051
Contact Information:0508 489 264
Manufacturer:Kendall, a division of Tyco Healthcare Group LP
Issue:Covidien has updated the instructions for use for polyurethane single lumen umbilical vessel catheters to include additional information on the care of the device. The updates were made to address the increasing use of disinfectants containing alcohol and/or acetone to clean vascular access devices. While the manufacturer recognizes that these materials are now required in many hospital protocols for routine cleaning of vascular access devices, the use of these disinfectants can cause damage to the polyurethane contained in the catheter and make it more susceptible to physical damage, which may ultimately lead to the need to replace the catheter.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/9/2013