Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15374
Brand Name:Trinity Biotech Captia Cytomegalovirus (CMV) IgG
Model:Product Code: 2325200
Affected:Lot Number: 557
Software version:
Recalling Organisation:Banksia Scientific Pty Ltd (NZ)
12 Brigade Street
Airport Oaks
Auckland
Contact Information:09 921 2626
Manufacturer:Trinity Biotech
Issue:Upon review of the QA release testing, the range that was calculated at the time of kit release was calculated per standard operating procedures and supports the Positive Control range that was assigned upon release of 2.2-4.1.
However, upon investigation of history (as a Positive Control in several kit lots) and additional testing it has been re-evaluated and determined that this range requires an adjustment to 2.2-4.6.
If you have used this kit lot to test and report patient results, and the Positive Control was within the assigned range (2.2-4.1), then the reported patient results were valid.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/9/2013