| Medsafe Reference: | 15359 |
| Brand Name: | Siemens Artis Imaging Systems |
| Model: | Catalogue Numbers: 10094141, 10094143 |
| Affected: | Version: VC1x software |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | Under certain conditions, x-ray release on Artis system version VC1x may get blocked. This potential malfunction will only occur if a) the system is ready for operation, and b) the C-arm was moved into the “Patient Transfer Position” for patient transfer (in this position x-ray is blocked), and c) at the same time the C-arm is in “Patient Transfer Position”, a hardware failure in the Image Acquisition System is occurring. Even if the C-arm is moved out of the “Patient Transfer Position” (the problem with the IAS was solved by the system itself), x-ray remains blocked. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/9/2013 |