Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15342 |
| Brand Name: | GE CardiacVX and CardiacVX Flow Analysis software |
| Model: | |
| Affected: | Version: 1.1.0 Ext.D |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems SCS |
| Issue: | CardiacVX and CardiacVX Flow report higher than actual flow values on phase-contrast images acquired with magnitude weighting mask enabled. Note that this issue can only occur on phase-contrast MR protocols with the magnitude weighting mask option enabled. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/9/2013 |