Detail of Recall

Type of Product:Device
Medsafe Reference:15331
Brand Name:Synthes Sternal ZipFix System
Model:Part Number: 03.501.080
Affected:Lot numbers prior to 8100630
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Issue:The Technique Guide includes a precaution that the trigger must be released before and during cutting and that no cutting should take place under tension, however when using the 1st generation Application Instrument for Sternal ZipFix™ it is possible to cut the implant while the tensioning trigger is being squeezed (known as being cut under tension).
NOTE: This issue applies to Application Instrument lots prior to 8100630 only (referred to as '1st generation' instruments).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/9/2013