Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15309
Brand Name:Biomerieux BacT/ALERT PF PLUS Culture Bottle
Model:Catalogue Number: 410853
Affected:Lot Number: 1033298
Software version:
Recalling Organisation:BioMerieux New Zealand Ltd
Suite 4, The Business Connection Office
2 Kalmia Street
Auckland 1544
Contact Information:0800 284 825
Manufacturer:BioMerieux Inc
Issue:BacT/ALERT® PF PLUS Culture Bottle, lot 1033298 yielded reflectance results above the stability specification at the 9-month timeframe. Bottles that have reflectance results near or greater than the initial value threshold specification when first loaded on the instrument could have an increased probability of false positive results.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/9/2013
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /