Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15289
Brand Name:Boston Scientific COGNIS Implantable Cardiac Defibrillator
Model:Models N106, P106
Affected:Devices manufactured prior to Dec-2009
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:Boston Scientific COGNIS Iimplantable Cardiac Defibrillator
Model:Models N107, P107
Affected:Devices manufactured prior to Dec-2009
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:Boston Scientific TELIGEN Implantable Cardiac Defibrillator
Model:Models E110 & F110
Affected:Devices manufactured prior to Dec-2009
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:Boston Scientific TELIGEN Implantable Cardiac Defibrillator
Model:Models E102 & F102
Affected:Devices manfactured prior to Dec-2009
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Issue:Boston Scientific has determined that the performance of an electrical component (a low voltage capacitor) in this subset of devices may be compromised over time, causing increased current drain that can lead to premature battery depletion. All cases to date have been detected by diagnostic tools within Boston Scientific's Safety Architecture before device function was compromised.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:30/8/2013