Detail of Recall

Type of Product:Device
Medsafe Reference:15287
Brand Name:Acclarent Inspira Air Balloon Dilation System
Model:Product Codes: BC0524A, BC0524AZ, BC0724A, BC0724AZ, BC1040A, BC1040AZ, BC1440A, BC1440AZ
Affected:All lots
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Acclarent Inc
Issue:Reports of difficulty deflating the balloon during a procedure have been received. In each of these reports, difficulty deflating occurred after the surgeon had pulled against resistance in response to the balloon moving distally during dilation. The force applied to the catheter stretched and narrowed the balloon catheter shaft, causing the balloon to be difficult or impossible to deflate. The potential for this phenomenon to occur exists in all lots and all sizes of the Inspira AIR Balloon Dilation System.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/9/2013