Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15287 |
| Brand Name: | Acclarent Inspira Air Balloon Dilation System |
| Model: | Product Codes: BC0524A, BC0524AZ, BC0724A, BC0724AZ, BC1040A, BC1040AZ, BC1440A, BC1440AZ |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Acclarent Inc |
| Issue: | Reports of difficulty deflating the balloon during a procedure have been received. In each of these reports, difficulty deflating occurred after the surgeon had pulled against resistance in response to the balloon moving distally during dilation. The force applied to the catheter stretched and narrowed the balloon catheter shaft, causing the balloon to be difficult or impossible to deflate. The potential for this phenomenon to occur exists in all lots and all sizes of the Inspira AIR Balloon Dilation System. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/9/2013 |